![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 418](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759418.webp)
System Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.256
E-6
Add/Modify External Power Meter
Selects and configures an external power meter.
The USB power meters supported at present are configured automatically, all input fields except
the optional Name field are disabled. A particular driver and serial number can be selected via
remote control.
Remote control:
SYSTem:COMMunicate:RDEVice:GENerator<gen_no>:DEFine
External Generators
Configures external generators with their connection type and device address.
Use of external generators, examples
An external generator can serve different purposes:
•
Extended measurement functionality: Each external generator represents an additional source
port. External generators increase the number of RF input signals for the DUT. A typical
example is a mixer measurement with a 2-port analyzer, where an external generator provides
the LO input signal.