R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.207
E-6
of the calibration sweep and the result can be monitored in the diagram. In case of an error (e.g. if the
measurement result shows that the calibration standard was not connected properly),
Abort Sweep
immediately terminates the sweep.
After completing the sweep the analyzer generates a short sound and a green checkmark appears in
the checkbox. Measurements can be repeated as often as desired. Newer results overwrite older
measurement data.
The trigger settings are ignored for calibration sweeps. Each calibration sweep is started without
delay (Free Run measurement). This also holds for power calibration sweeps.
Checks for the calibration sweep
If the sweep range of the active channel exceeds the validity range of the standard model (defined by
Min. Freq.
and
Max. Freq.
in the
Add/Modify Standard
dialog) the analyzer displays a notice box
(confirm with
OK).
The
Measure Standards
dialog provides further controls:
•
Keep Measurement Data for >Repeat Cal<
causes the raw measurement data of the
standards to be stored after the calibration is completed. This enables the
Repeat Previous
Cal...
command, which can be used to optimize a previous calibration without repeating the
measurement of all standards. If
Keep Measurement Data...
is not active, then the raw
measurement data of the standards is deleted and the analyzer only stores the system error
correction data. Deleting the raw data saves disk space.
•
Apply
is enabled as soon as data have been acquired for all standards. The button starts the
calculation of the system error correction data and closes the calibration wizard. The current
instrument settings are stored with the correction data. To avoid incompatibilities, older system
error correction data is deleted unless it has been transferred into a
Cal Pool
using the
Calibration Manager.
The
Keep Measurement Data for >Repeat Cal<
setting is valid for the current calibration only. To
activate this function in general, use the parameter in the
User Interface
tab of the
Sytem Config.
dialog
(menu
System – System Config
).
Checks during the calculation of correction data
Incompatibilities between the selected calibration type, the standards and the channel settings may
cause the calibration to be inaccurate
.
The analyzer auto-detects potential sources of errors and
displays appropriate, self-explanatory notice boxes.
Remote control:
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe
[SENSe<Ch>:]CORRection:COLLect[:AQUire]:SELected
[SENSe<Ch>:]CORRection:COLLect:SAVE
[SENSe<Ch>:]CORRection:COLLect:DELete ["<cal_name>"]
Repeat Previous Cal
Reopens the
Measure Standards
dialog of the calibration wizard to repeat and optimize the previous
calibration without necessarily repeating the measurement of all standards.
Repeat Previous Cal
is active only if
Keep Measurement Data for >Repeat Cal<
in the
Measure
Standards
dialog was enabled when the last calibration for the active channel was successfully
completed.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......