Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.144
E-6
The sweep segment file actually contains more columns listing all channel settings of the
individual sweep segments. The headings of the additional columns read:
Source Power [dBm]; IF Bandwidth [Hz]; enIF Selectivity; en IF Sideband; Meas Delay [
µ
s];
boSweep Time Auto;
Number of Points
Sets the total number of measurement points per sweep. The minimum number of points is 2.
Together with the sweep range defined with the
Stimulus
settings, this parameter defines the grid of
sweep points. The sweep points are equidistantly distributed over the entire sweep range: The step
width between two consecutive sweep points is constant on a linear scale (sweep types
Lin. Frequency,
Time
and
CW Mode)
or on a logarithmic scale (sweep types
Log. Frequency
and
Power)
.
In
Segmented Frequency
sweeps, the number of points can be set independently for each segment;
see
Define Segments
.
As an alternative to the
Number of Points,
the
Stimulus Step Size
can be set.
Measurement time and screen resolution
A large number of points improves the resolution of the trace but increases the measurement time.
The overall measurement time is composed of a hardware settling time at the beginning of the sweep
plus the sum of the measurement times at each individual sweep point. This implies that the
measurement time increases roughly linearly with the number of points.
After changing the channel settings or selecting another measured quantity, the analyzer needs
some time to initialize the new sweep. This preparation period increases with the number of
points and the number of partial measurements involved. It is visualized by a Preparing Sweep
symbol in the status bar:
All analyzer settings can still be changed during sweep initialization. If necessary, the analyzer
terminates the current initialization and starts a new preparation period.
Remote control:
[SENSe<Ch>:]SWEep:POINts
Frequency Step Size
Sets the distance between two consecutive frequency sweep points.
The step size is an alternative to the
Number of Points
setting:
•
If the sweep range is defined by means of the
Start
and
Stop
variables, both the
Stop
value and
the
Number of Points
can vary as the
Stimulus Step Size
is changed. The
Stop
value is
changed as little as possible so that the condition
Stimulus Step Size = (Stop – Start) / (Number
of Points – 1)
can be fulfilled
.
Changing the
Start and Stop
values
modifies the
Stimulus Step
Size.
•
If the sweep range is defined by means of the
Center
and
Span
variables, both the
Span
value
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......