![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 223](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759223.webp)
R&S
®
ZVA/ZVB/ZVT
Trace
Menu
1145.1084.12 4.61
E-6
factor of 2. This increased computation time is usually negligible compared to the sweep times of the
analyzer.
The following properties of the Chirp z-transformation are relevant for the analyzer settings:
•
The frequency points must be equidistant. If a
Log Frequency
or
Segmented Frequency
sweep
is active, the measurement points are interpolated.
•
The time domain response is repeated after a time interval which is equal to t = 1/ f, where f
is the spacing between two consecutive sweep points in the frequency domain. For a sweep
span of 4 GHz and 201 equidistant sweep points, f = 4 GHz/200 = 2 * 10
7
Hz, so that t = 50
ns. t is termed measurement range (in time domain) or unambiguous range.
Additional constraints apply if the selected Chirp z-transformation is a lowpass transformation.
Remote control:
CALCulate<Ch/Tr>:TRANsform:TIME:STATe ON
Time Domain Stimulus Axis
Opens a submenu to define the stimulus axis range for time domain representation.
•
Start Time Domain
is the lowest displayed time and corresponds to the left edge of the
Cartesian diagram.
•
Stop Time Domain
is the highest displayed time and corresponds to the right edge of the
Cartesian diagram.
•
Center Time Domain
corresponds to the center of the Cartesian diagram, i.e. (Start + Stop)/2.
•
Span Time Domain
corresponds to the diagram width, i.e. (Stop – Start).
•
Time
and
Distance
switch over between the x-axis scaling in time units or distance units.
•
Set Frequencies
opens the
Channel
–
Stimulus
menu in order to change the stimulus axis in
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......