![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 245](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759245.webp)
R&S
®
ZVA/ZVB/ZVT
Trace
Menu
1145.1084.12 4.83
E-6
•
Freq Re(S11) Im(S11) Re(S21) Im(S21) Re(S12) Im(S12) Re(S22) Im(S22)
•
(all values arranged in 1 line)
•
3-port files (*.s3p)
•
Freq Re(S11) Im(S11) Re(S12) Im(S12) Re(S13) Im(S13)
•
Re(S21) Im(S21) Re(S22) Im(S22) Re(S23) Im(S23)
•
Re(S31) Im(S31) Re(S32) Im(S32) Re(S33) Im(S33)
•
(values arranged in 3 lines)
•
4-port files (*.s4p)
•
Freq Re(S11) Im(S11) Re(S12) Im(S12) Re(S13) Im(S13) Re(S14) Im(S14)
•
Re(S21) Im(S21) Re(S22) Im(S22) Re(S23) Im(S23) Re(S24) Im(S24)
•
Re(S31) Im(S31) Re(S32) Im(S32) Re(S33) Im(S33) Re(S34) Im(S34)
•
Re(S41) Im(S41) Re(S42) Im(S42) Re(S43) Im(S43) Re(S44) Im(S44)
•
(values arranged in 4 lines)
The stimulus frequencies are arranged in ascending order. If a
lin. Mag-Phase (MA)
or
dB Mag-Phase
(DB)
data format is selected the real and imaginary S-parameter values Re(S
ij
), Im(S
ij
) are replaced by
lin Mag(S
ij
), phase(S
ij
) or dB Mag(S
ij
), phase(S
ij
), respectively.
ASCII (*.csv) files
An ASCII file contains a header and the actual trace data:
freq;reTrc1_S21;imTrc1_S21;reMem2[Trc1]_S21;imMem2[Trc1]_S21;
300000.000000;0.000000;0.000000;0.000000;0.000000;
40499497.487437;0.000000;0.000000;0.000000;0.000000;
80698994.974874;0.494927;-0.065174;0.500833;-0.074866;
120898492.462312;0.497959;-0.111724;0.488029;-0.107375;
...
The header consists of the following data elements:
<Stimulus>
stimulus variable: freq for Frequency sweep, power for Power sweep, time for Time
sweep, trigger for CW Mode sweep.
<reTrace1>
first response value of first trace: re<Trace_Name>, mag<Trace_Name> or
db<Trace_Name> for output format Re/Im, lin. Mag-Phase or dB Mag-Phase, respectively. The data
format for export files can be selected in the Export Data dialog.
<imTrace1>
<reTrace2>
<imTrace2>...
second response value of second trace. first response value of second trace.
HZ / KHZ / MHZ / GHZ allowed for imported files. The analyzer alway uses HZ for exported
data.
second response value of first trace: im<Trace_Name> for output format Re/Im,
ang<Trace_Name> for output formats lin. Mag-Phase or dB Mag-Phase. The data format for export
files can be selected in the Export Data dialog.
The trace data is arranged as described in the header. Different values are separated by semicolons. A
semicolon is also inserted before the end of each line.
The stimulus values are arranged in ascending order.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......