![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 531](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759531.webp)
R&S
®
ZVA/ZVB/ZVT
CALCulate
1145.1084.12 6.69
E-6
CALCulate<Ch/Tr>:LIMit:RDOMain:SPACing LINear | LOGarithmic | dB | SIC
Deletes the existing limit line and (re-)defines the physical units of the response values of the limit line.
The units of the stimulus values are defined via
CALCulate<Ch/Tr>:LIMit:CONTrol:DOMain.
This command is complemented by
CALCulate<Ch/Tr>:LIMit:RDOMain:COMPlex
and
CALCulate<Ch/Tr>:LIMit:RDOMain:FORMat
.
<Ch/Tr>
Channel number used to identify the active trace.
Parameters
Keyword for the physical unit of the response values.
Range [def. unit]
The analyzer uses dB units, irrespective of the parameter
selected.
[–]
*RST value
–
SCPI, Command
Types
Device-specific, no query.
CALCulate<Ch/Tr>:LIMit:SEGMent<Seg>:AMPlitude:STARt <numeric_value>
Changes the start response value (i.e. the response value assigned to the start stimulus value) of a limit
line segment. A segment must be created first to enable this command (e.g
CALC:LIM:DATA
).
To define the response values of several limit line segments with a single command, use
CALCulate<Ch/Tr>:LIMit:LOWer[:DATA]
or
CALCulate<Ch/Tr>:LIMit:UPPer[:DATA]
.
<Ch/Tr>
Channel number used to identify the active trace.
<Seg>
Segment number
<numeric_value>
Response value
Range [def. unit]
Almost no restriction for limit segments; see
Rules for Limit Line
Definition.
[dB]
*RST value
–
The default response values of a segment that is created by
defining its stimulus values only (e.g. by means of
CALCulate<Ch/Tr>:LIMit:CONTrol[:DATA]
), is –20 dB.
SCPI, Command
Types
Device-specific, with query.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......