Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.120
E-6
Stimulus and Sweep Types
The function and input values of the
Start, Stop, Center
and
Span
softkeys depend on the sweep type:
Sweep type
Start [dimension]
Stop [dimension]
Center [dimension]
Span [dimension]
Equivalent hotkey
(press twice)
(press twice)
Lin. Frequency
Log. Frequency
Segmented Frequency
Start frequency [Hz]
Stop frequency [Hz]
Center frequency [Hz]
Frequency span [Hz]
Power
Start power [dBm]
Stop power [dBm]
–
–
Time
–
Total sweep time [s]
–
–
CW Mode
–
Number of points [1]
–
–
The ranges of numerical values must be compatible with the instrument model.
If a frequency sweep is selected, then the
START CENTER
and
STOP SPAN
are both toggle
keys. In
Time
or
CW Mode
sweep mode the
Start, Center
and
Stop
softkeys are inactive, still you can
use the
START CENTER
hardkey to select the fixed CW frequency.
The conditions for the stimulus range depend on the sweep type:
•
Lin. Frequency/Log. Frequency/Segmented Frequency
The frequency range depends on the instrument type; see SCPI command
reference. The
Stop
frequency must be greater than the
Start
frequency; the
Span
must be 1 Hz. If a
Stop
frequency smaller than the current
Start
frequency is set, then the
Start
frequency is adjusted and vice versa).
A sweep must contain at least two different sweep points.
•
Power
Start
and
Stop
power are entered in absolute units (dBm) and over a wide
range. The
Center
and
Span
softkeys are not active.
Start
and
Stop
power must be different, the
Stop
power must be larger than
the
Start
power. If a
Stop
power smaller than the
Start
power isset, then the
Start
power is adjusted automatically and vice versa.
The power corresponds to the actual output power at the test port that supplies
the stimulus for the active channel or at the calibrated reference plane.
•
Time
The
Stop
value represents the total sweep time.
Start
and the other
parameters are not active as the sweep time is relative to the beginning of the
sweep which starts immediately after the entry of the
Stop
value.
The sweep time is entered in seconds and must be positive; see also Time
sweep.
•
CW Mode
The
Stop
value represents the number of trigger events during the
measurement. Each trigger event triggers one measurement point.
Start
and
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......