![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 684](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759684.webp)
SENSe
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.222
E-6
*RST value
The default frequency/frequency range corresponds to the sweep
range or CW frequency of the analyzer, i.e.
<numerator> = <denominator> = 1
<offset> = 0 Hz
Sweep type
Parameters
SWEep – A frequency sweep is active, the command defines a
frequency range.
CW | FIXed – A power, time or
CW Mode
sweeps is active, the
command defines a fixed frequency.
*RST value
SWEep
SCPI, Command
Types
Device-specific, command or query.
Example:
*RST; FREQ:STAR 1E+9; STOP 1.1E+9
Reset the analyzer (activating a frequency sweep) and set the
sweep range between 1 GHz and 1.1 GHz.
SENS:FREQ:CONV:ARB 2, 1, 1E+9, SWE
Convert the receiver frequency to the range between 3 GHz and
3.2 GHz.
[SENSe<Ch>:]FREQuency:CONVersion:AWReceiver[:STATe] <Boolean>
Qualifies whether a waves are measured at the source or at the receiver frequency. The setting is
relevant for all frequency-converting measurements.
<Ch>
Channel number.
<Boolean>
OFF - Measurement at source frequency
ON - Measurement at receiver frequency
*RST value
OFF
SCPI, Command Types
Device-specific, command or query.
Example:
FREQ:CONV:AWR ON
Measure a waves at the receiver frequency.
[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:ORDer <order>
Selects the order of the harmonic measured if a harmonic measurement is enabled
(
[SENSe<Ch>:]FREQuency:CONVersion HARMonic
).
<Ch>
Channel number.
<order>
Order of the harmonic, integer multiple of the fundamental
frequency measured.
Range [def. unit]
2 to 100 [–].
*RST value
2
SCPI, Command
Types
Device-specific, command or query.
Example:
See [SENSe<Ch>:]FREQuency:CONVersion
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......