Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM OFF
Disable the measurement in the created sweep segment.
SEGM:COUN?
Query the number of segments.
[SENSe<Ch>:]SEGMent<Seg>:DEFine
<Start>,<Stop>,<Points>,<Power>,<Time>|<Point Delay>,<Unused>,<Meas.
Bandwidth>[,<LO>,<Selectivity>]
Re-defines a sweep segment with specific channel settings
(Insert New Segment)
. The segment
replaces an existing segment <Seg> in the segment list. The modified segment must not overlap with
any of the existing segments.
Entry of the first seven numeric parameters is mandatory; no default values are provided. All settings
except <LO> can be changed for existing segments using other commands of the
[SENSe<Ch>:]SEGMent<Seg>...
subsystem.
Note:
Use
[SENSe<Ch>:]SEGMent<Seg>:Add
to create a segment with default channel settings.
Use
[SENSe<Ch>:]SEGMent<Seg>:INSert
(no query) to insert a new segment into the
current segment list.
<Ch>
Channel number.
<Seg>
Sweep segment number. Segment numbers must be sequential. The
specified segment number must be smaller or equal to the number of
existing segments plus 1.
If segment number <Seg> already exists, it is replaced by the new
segment.
<Start>, <Stop>
Start and stop frequency of the segment. See
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP.
Range [def. unit]
A new segment must not overlap with any of the existing segments.
Besides the frequency range depends on the analyzer model. [Hz]
<Points>
Number of sweep points in the segment. See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts.
Range [def. unit]
1 to 2147483647. [1] 1 is allowed if start and stop frequencies are
equal.
<Power>
Internal source power in the segment. See
[SENSe<Ch>:]SEGMent<Seg>:POWer.
Range [def. unit]
–40 dBm to +10 dBm. The exact range depends on the analyzer
model; refer to the data sheet [dBm]. UP and DOWN
increment/decrement the source power in 1-dB steps.
<Time>
Duration of the sweep in the segment. See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME
. In the setting
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect DWELl,
this
parameter is replaced by <Point Delay>.
Range [def. unit]
Depending on other channel settings [s].
AUTO
activates automatic
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......