Optimization
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 9.6
E-4
sweep is performed at maximum speed without delay.
Avoid delay times between the end of the measurement sequence and the next trigger event:
2. If you don't need a special trigger mode, click
Channel – Sweep – Trigger
and make sure that
Free Run
is selected
.
Avoid delay times between adjacent frequency points.
3. Click
Channel – Mode
and select
Alternating Sweeps
if the settling time between adjacent
frequency points is larger than the settling time between the partial measurements.
To find the appropriate noise reduction settings...
Use the fastest IF filter that will produce acceptable dynamic range:
1. Press
the
POWER BW AVG
key in the
Channel
block or click
Channel – Power Bandwidth
Average
to access the corresponding
submenu.
2. Select the widest bandwidth compatible with your measurement task
.
3. If possible, click
Fine Adjust
to further increase the bandwidth. Avoid selecting
High Selectivity
unless it is required for your measurement.
Reduce the number of repeated measurements made to obtain an averaged result. Averaging does not
slow down the measurement, however, several consecutive sweeps must be performed until an
average result is available.
1. If averaging is not necessary, open the
Channel – Power Bandwidth Average
menu
and
make
sure that
Average On
is not active.
2. Alternatively,
click
Average Factor
and reduce the factor to a minimum.
Calibrating a Measurement Channel
Calibration is the process of eliminating systematic, reproducible errors from the measurement results
(system error correction). A measurement channel can be calibrated either manually or automatically
using the Calibration Unit (option R&S ZV-Z51):
To select and perform a manual calibration...
In the following example a 2-port TOSM calibration is performed manually using the Calibration Wizard.
Other calibrations can be performed in an analogous way.
1. Perform all channel settings for your measurement.
2. Click
Channel – Calibration – Start Cal – Two Port P1 P2 – Manual...
to open the calibration
wizard for the desired calibration type.
3. In
the
Select Physical Port Connector(s)
dialog, select the connector type that you use at port 1
and 2 (both connector types must be equal) and the calibration kit. Click
Next >
to access the
Measure Standards
dialog
.
4. Connect an open standard to port 1 and click the corresponding box in the
Measured
Standards
list to initiate the calibration sweep and acquire the calibration data for this standard.
5. Repeat the last step for a short and a match connected to port 1, an open, short and match
connected to port 2, and for a through between both ports.
6. Click
Apply
to
start the calculation of the system error correction data, apply them to the current
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......