R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.229
E-6
The default delay is 0 s, the default step width is 1 ns, corresponding to a step width of 299.792
mm for the electrical length. The relations hold for one-port and 2-port standards.
•
Z
0
is the
Characteristic Impedance
of thestandard. If the standard is terminated with Z
0
, then its
input impedance is also equal to Z
0
.
Z
0
is not necessarily equal to the reference impedance of
the system (depending on the
Connector Type)
or the terminal impedance of the standard. The
characteristic impedance of the standard is only used in the context of calibration.
The default characteristic impedance is equal to the reference impedance of the system.
•
The
Loss
is the energy loss along the transmission line due to the skin effect. For resistive lines
and at RF frequencies the loss is approximately proportional to the square root of the
frequency.
In Agilent mode the
Offset Loss
is expressed in units of /s at a frequency of 1 GHz. The
following formula holds:
To determine an offset loss value experimentally, measure the delay in seconds and the loss in
dB at 1 GHz and use the formula above.
The default
Loss
or
Offset Loss
is zero.
The impedance for waveguides is frequency-dependent. If a waveguide line type is selected in
the Offset Model dialog, the Char. Impedance field is disabled and indicates "varies" instead of a
definite impedance value. Moreover no Loss or Offset Loss can be set.
Offset parameters and standard types
Offset parameters are used to describe all types of standards except the
Sliding Match
and the
Attenuation.
•
The
Sliding Match
is a one-port standard with variable load parameters (sliding load) and
unspecified length. The reference impedance is fixed and equal to the characteristic impedance
of the connector type
.
No load and offset parameters need to be set.
•
The
Attenuation
is a two-port standard which is fully matched in both directions (the reflection
factor at both ports is zero). No load and offset parameters need to be set.
Remote control:
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type>
Modify Load
Specifies the load parameters for a particular calibration standard describing its terminal impedance.
This dialog is opened from the
Add or View / Modify Standard...
dialog (
Modify Load...
button).
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......