![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 262](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759262.webp)
Trace Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.100
E-6
•
A bandpass region contains a local maximum around which the magnitude of the trace falls off
by more than a specified
x dB Bandwidth
.
•
A bandstop region contains a local minimum around which the magnitude of the trace increases
by more than a specified
x dB Bandwidth
.
The analyzer locates bandpass and bandstop regions and determines their position
(Center
frequency)
and shape
(Bandwidth, LBE, UBE
, Quality factor
Q
; see
Show Results).
For a meaningful definition of
the
x dB Bandwidth
criterion, the trace format must be
dB Mag.
•
Bandpass Search Ref to Max
activates the search for a bandpass region in the active trace,
starting at the absolute maximum of the active trace within the search range.
•
Bandstop Search
activates the search for a bandstop region in the active trace, starting at the
absolute minimum of the active trace within the search range.
•
Bandfilter Tracking
causes the bandfilter search to be repeated after each sweep.
•
x dB Bandwidth...
sets the level defining the filter bandwidth.
•
Search Range...
confines the search to a subrange of the sweep.
•
Search Result Off
hides the display of the bandfilter parameters in the diagram area.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......