![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 764](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759764.webp)
TRIGger<Ch>...
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.302
E-6
MANual
Trigger signal generated by pressing the
Manual Trigger
softkey.
*RST value
IMMediate
SCPI,
Command
Types
Confirmed, command or query.
Example:
TRIG:SOUR MAN
Activate manual trigger mode. The analyzer starts the next
sweep when the
Manual Trigger
softkey is pressed.
TRIGger<Ch>[:SEQuence]:TIMer <numeric_value>
Sets the period of the internal periodic signal that can be used as a trigger source
(
TRIGger<Ch>[:SEQuence]:SOURce TIMer
).
<Ch>
Channel number. If unspecified the numeric suffix is set to 1.
<numeric_value>
Timer period
Range [def. unit]
1E-005 s to 13680
s [s]. UP and DOWN increment/decrement
the timer period in 10 ms steps.
*RST value
1 s
SCPI, Command
Types
Confirmed, command or query
Example:
TRIG:SOUR TIM
Activate internal timer as trigger source.
TRIG:TIM 0.1
Set a timer period of 100 ms.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......