Standard Type
Characteristics
Ideal Standard
(Dummy
Parameters)
Offset Model
Load
Model
Match
Matched broadband termination (one-port)
Z
0
(reference
impedance of the
connector type)
Sliding match
One-port standard with variable load
parameters (sliding load), unspecified
length and zero loss. The reference
impedance is fixed and equal to the
reference impedance of the connector
type
.
The sliding match is measured at 3 to
7 positions of the load for each calibrated
port. This improves effective directivity and
thus the calibration accuracy, in particular
in the high-frequency range where the
Match
standards can cause reflections.
–
–
–
Reflect
Unknown mismatched standard (one-port)
Through Through-connection
with
minimum
loss
(two-port)
–
–
Line
Line for TRL calibration with minimum loss
(two-port)
–
–
Attenuation
Fully matched standard in both directions
(two-port; the reflection factor at both ports
is zero).
–
–
–
Symm. network
Unknown mismatched reflection-symmetric
standard (two-port)
–
Remote control:
For an overview of standard parameters see also
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type>
Channel - Offset
The
Offset
menu
defines a length offset for the test ports. The offset complements the system error
correction, compensating for the known length of a (non-dispersive) transmission line between the
calibrated reference plane and the DUT.
Offset parameters: Definition
The
Delay
is the propagation time of a wave traveling through the transmission line. The
Electrical Length
is equal to the
Delay
times the speed of light in the vacuum and is a measure
for the length of transmission line between the standard and the actual calibration plane. For a
line with permittivity
r
and mechanical length L
mech
the delay and the electrical length are
calculated as follows:
The use of an
Electrical Length, Mechanical Length
or
Delay
is therefore equivalent; the
analyzer automatically calculates all three parameters when a single value is entered.
For a non-dispersive DUT, the delay defined above is constant over the considered frequency
range and equal to the negative derivative of the phase response with respect to the frequency
(see mathematical relations). The offset parameters compensate for a constant delay, which is
equivalent to a linear phase response.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......