![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 604](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759604.webp)
DISPlay...
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.142
E-6
Example:
*RST; :DISP:WIND:TRAC:X:OFFS 1MHZ; :DISP:WIND:TRAC:Y:OFFS 10
Create the default trace and shift it horizontally by 1 MHz, vertically
by 10 dB.
DISP:WIND:TRAC:Y:OFFS?
Query all response offset values. The response is
10,0,0,0.
DISPlay:WINDow<Wnd>:TRACe<WndTr>:Y[:SCALe]:AUTO ONCE
Displays the entire trace <WndTr> in the diagram area, leaving an appropriate display margin.
<Wnd>
Number of an existing diagram area (defined by means of
DISPlay:WINDow<Wnd>:STATe ON
).
<WndTr>
Existing trace number, assigned by means of
DISPlay:WINDow<Wnd>:TRACe<WndTr>:FEED
.
ONCE
Activate the autoscale function.
*RST value
–
SCPI,
Command
Types
Confirmed, no query.
Example:
*RST; DISP:WIND:TRAC:Y:PDIV?; RLEV?
Query the value between two grid graticules and the reference value
for the default trace. The response is
10;0.
DISP:WIND:TRAC:Y:AUTO ONCE; PDIV?; RLEV?
Autoscale the default trace and query the scaling parameters again. In
general both values have changed.
DISPlay:WINDow<Wnd>:TRACe<WndTr>:Y[:SCALe]:PDIVision
<numeric_value>
Sets the value between two grid graticules (value “per division”) for a particular displayed trace. When a
new PDIVision value is entered, the current RLEVel is kept the same, while the top and bottom scaling
is adjusted for the new PDIVision value.
<Wnd>
Number of an existing diagram area (defined by means of
DISPlay:WINDow<Wnd>:STATe ON
).
<WndTr>
Existing trace number, assigned by means of
DISPlay:WINDow<Wnd>:TRACe<WndTr>:FEED
.
<numeric_value>
Value and unit for the vertical diagram divisions.
Range [def. unit]
Range and unit depend on the measured quantity, see
Units for
DISPlay... Commands
.
*RST value
Depending on the measured quantity. The default reference level
for an S-parameter displayed in a
dB Mag
diagram is 10 dB.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......