R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.154
E-6
DISPlay:WINDow<Wnd>:TRACe<WndTr>:Y[:SCALe]:PDIVision
<numeric_value>
Sets the value between two grid graticules (value “per division”) for a particular displayed trace. When a
new PDIVision value is entered, the current RLEVel is kept the same, while the top and bottom scaling
is adjusted for the new PDIVision value.
<Wnd>
Number of an existing diagram area (defined by means of
DISPlay:WINDow<Wnd>:STATe ON
).
<WndTr>
Existing trace number, assigned by means of
DISPlay:WINDow<Wnd>:TRACe<WndTr>:FEED
.
<numeric_value>
Value and unit for the vertical diagram divisions.
Range [def. unit]
Range and unit depend on the measured quantity, see
Units for
DISPlay... Commands
.
*RST value
Depending on the measured quantity. The default reference level
for an S-parameter displayed in a
dB Mag
diagram is 10 dB.
SCPI, Command
Types
Confirmed, with query.
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named
Ch4Tr1
to measure the input
reflection coefficient S11.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC9:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2, assigning the
trace number 9 to it.
DISP:WIND2:TRAC9:Y:RLEV -10
Change the reference level to –10 dB.
DISPlay:WINDow<Wnd>:TRACe<WndTr>:Y[:SCALe]:RLEVel <numeric_value>
Sets the reference level (or reference value) for a particular displayed trace. Setting a new reference
level does not affect the value of PDIVision.
<Wnd>
Number of an existing diagram area (defined by means of
DISPlay:WINDow<Wnd>:STATe ON
).
<WndTr>
Existing trace number, assigned by means of
DISPlay:WINDow<Wnd>:TRACe<WndTr>:FEED
.
<numeric_value>
Value and unit for the reference level (or reference value, if the
trace does not show a level).
Range [def. unit]
Range and unit depend on the measured quantity, see
Units for
DISPlay... Commands
.
*RST value
Depending on the measured quantity. The default reference level
for an S-parameter displayed in a
dB Mag
diagram is 0 dB.
SCPI, Command
Confirmed, with query.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......