![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 523](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759523.webp)
R&S
®
ZVA/ZVB/ZVT
CALCulate
1145.1084.12 6.61
E-6
CALCulate<Ch/Tr>:LIMit...
This subsystem defines the limit lines and controls the limit check.
CALCulate<Ch/Tr>:LIMit:CONTrol[:DATA]
<numeric_value>,<numeric_value>{,<numeric_value>,<numeric_value>}
Defines the stimulus values of the limit line and/or creates new limit line segments.
Rules for creating segments
The following rules apply to an active trace with n existing limit line segments:
An odd number of values is rejected; an error message –109,"Missing parameter..." is
generated.
An even number of 2*k values updates or generates k limit line segments.
For n > k the stimulus values of all existing limit line segments no. 1 to k are updated, the
existing limit line segments no. k+1, ..., n are deleted.
For n < k the stimulus values of the limit line segments no. 1 to n are updated, the limit line
segments n+1, ,..., k are generated with default response values (see
CALCulate<Ch/Tr>:LIMit:UPPer[:DATA],
CALCulate<Ch/Tr>:LIMit:LOWer[:DATA]
).
The generated segments are upper or lower limit line segments, depending on the
CALCulate<Ch/Tr>:LIMit:SEGMent<Seg>:TYPE setting.
CALCulate<Ch/Tr>:LIMit:CONTrol[:DATA]
does not overwrite the type setting.
To define additional new limit line segments without overwriting the old segments use
CALCulate<Ch/Tr>:LIMit:DATA
.
<Ch/Tr>
Channel number used to identify the active trace.
<numeric_value>
Pair(s) of frequency, power or time values, to be defined in
accordance with the domain setting
(
CALCulate<Ch/Tr>:LIMit:CONTrol:DOMain
).
Range [def. unit]
Almost no restriction for limit segments; see
Rules for Limit Line
Definition.
[Hz for frequency sweeps, dBm for power sweeps, s for
time sweeps]
*RST value
–
A segment that is created implicitly, e.g. by means of
CALCulate<Ch/Tr>:LIMit:UPPer[:DATA]
or
CALCulate<Ch/Tr>:LIMit:LOWer[:DATA,]
, covers the
maximum sweep range of the analyzer.
SCPI, Command
Types
Confirmed, with query.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......