![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 256](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759256.webp)
Trace Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.94
E-6
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:EXECute MAXimum
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:RESult?
Min Search
Sets the active marker to the
absolute minimum in the search range, i.e. to the smallest of all response
values. If a complex trace format (e.g. a polar diagram) is active, the marker is set to the measurement
point with the minimum magnitude.
By default the search range coincides with the sweep range. If the active trace contains no markers, a
marker
Mkr 1
is created to indicate the search result.
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:EXECute MINimum
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:RESult?
Next Peak
Sets the active marker to the next maximum or minimum in the search range, depending on the current
search criterion.
•
If a
Max Search
or a
Bandpass Search
is active, then the marker is set to the next maximum.
The next maximum is the maximum with the largest response value that is below the current
marker response value.
•
If a
Min Search
or a
Bandstop Search
is active, then the marker is set to the next minimum. The
next minimum is the minimum with the smallest response value that is above the current
marker response value.
By default the search range coincides with the sweep range. If the active trace contains no markers, a
marker
Mkr 1
is created to indicate the search result.
Next Peak
is disabled while a
Target Search
is
active.
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:EXECute
NPEak
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:RESult?
Peak Search >
Sets the active marker to the
next peak with higher stimulus value. If the active trace contains no
markers, a marker
Mkr 1
created to indicate the search result. The peak criteria are defined via Define
Peak.
Remote control:
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:EXECute RPEak
CALCulate<Ch/Tr>:MARKer<Mk>:FUNCtion:RESult?
< Peak Search
Sets the active marker to the
next peak with lower stimulus value. If the active trace contains no
markers, a marker
Mkr 1
created to indicate the search result. The peak criteria are defined via Define
Peak.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......