![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 530](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759530.webp)
CALCulate
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.68
E-6
CALCulate<Ch/Tr>:LIMit:RDOMain:COMPlex S | SINV | Y | Z | YREL
Deletes the existing limit line and (re-)defines the physical units of the response values of the limit line.
The units of the stimulus values are defined via
CALCulate<Ch/Tr>:LIMit:CONTrol:DOMain.
This command is complemented by
CALCulate<Ch/Tr>:LIMit:RDOMain:FORMat
and
CALCulate<Ch/Tr>:LIMit:RDOMain:SPACing
.
<Ch/Tr>
Channel number used to identify the active trace.
Parameters
Keyword for the physical unit of the response values.
Range [def. unit]
The parameters form four groups:
S and SINV select relative units (dB) for the limit line.
Y selects admittance units (S/Siemens).
Z selects impedance units (
Ω
).
YREL select dimensionless numbers (U).
[–]
*RST value
–
SCPI, Command Types
Device-specific, no query.
CALCulate<Ch/Tr>:LIMit:RDOMain:FORMat COMPlex | MAGNitude | PHASe |
REAL | IMAGinary | SWR | GDELay | L | C
Deletes the existing limit line and (re-)defines the physical units of the response values of the limit line.
The units of the stimulus values are defined via
CALCulate<Ch/Tr>:LIMit:CONTrol:DOMain.
This command is complemented by
CALCulate<Ch/Tr>:LIMit:RDOMain:COMPlex
and
CALCulate<Ch/Tr>:LIMit:RDOMain:SPACing
.
<Ch/Tr>
Channel number used to identify the active trace.
Parameters
Keyword for the physical unit of the response values.
Range [def. unit]
The parameters form four groups:
COMPlex, REAL, IMAGinary, and SWR select
dimensionless numbers (U) for the limit line.
MAGNitude selects relative units (dB).
PHASe selects phase units (deg).
GDELay selects time units (s).
L selects inductance units (H/Henry).
C selects capacitance units (F/Farad).
[–]
*RST value
–
SCPI, Command
Types
Device-specific, with query.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......