Optimization
1145.1084.12 9.5
E-4
To Set up a CW Mode Sweep...
1. Press
the
Sweep
key in the
Channel
keypad or click
Channel - Sweep
to access the
Sweep
submenu.
2. Click
Sweep Type - CW Mode.
3. In
the
CW Mode
entry bar opened, enter the fixed stimulus frequency
(CW Frequency)
and the
internal generator power
(Source Power)
for the sweep.
4. Click
Channel - Stimulus
to access the submenu defining the sweep range.
5. To vary the number of measurement points (implying a longer sweep time), use
Stop.
If the active diagram is a Cartesian diagram, the x-axis is linearly scaled and labeled with the current
numbers of the trigger events (measurement points).
Optimization
Optimizing the Measurement Speed
Minimizing the measurement time per sweep contributes to a high measurement throughputs. The
following parameters have an influence on the sweep time:
To find the best set of sweep points...
Limit the measurement to the stimulus range that you need for your device:
1. Click
Channel - Stimulus
to access the submenu defining the sweep range.
2. To define a sweep range confined by a start and a stop frequency, use
Start
and
Stop
.
Alternatively, to define a sweep range of definite width around a known center frequency (e.g.
an expected peak), use
Center
and
Span
. For power and time sweeps, proceed as described in
Setting up a Sweep.
3. Alternatively, use the marker functions to define the sweep range.
Use the minimum number of points that will produce an acceptable resolution:
1. Click
Channel – Sweep – Number of Points
and reduce the number of sweep points.
Use the segmented sweep to focus on different subranges of the sweep, using optimized channel
settings for each subrange:
1. Set up a segmented frequency sweep as described in Segmented Frequency.
2. Optimize the number of points and the noise reduction settings for each individual sweep
segment.
To obtain the fastest measurement sequence...
Avoid delay times between the different partial measurements:
1. Click
Channel – Sweep – Sweep Time
and check
Auto
.
The analyzer determines the minimum sweep time according to the current channel settings. The
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......