Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.208
E-6
Additional conditions for repeating a calibration
Changing the channel settings can cause incompatibilities between the current instrument state and the
previous calibration. In particular the following conditions must be met:
•
The configuration of all ports needed must be correct (see
Select Connector
–
Physical Port
Number #)
.
•
The
Sweep Type
must be unchanged.
If incompatible settings occur the analyzer displays an error message and does not repeat the
calibration.
Recall Last Cal Set
•
Loads and activates the setup for which the last calibration was performed. If the last
calibrated setup is already active, nothing is changed.
•
The calibrated setups are automatically stored in the
C:\Rohde&Schwarz\Nwa\Calibration\RecallSets
directory. A message box pops up if the
directory is empty, e.g. because no calibration was performed yet.
Correction Off
Activates or deactivates the system error correction in the active channel.
Correction Off
is active only if
a valid system error correction is assigned to the active channel; see
Cal State
in the
Calibration
Manager.
A red label Cal Off ! appears behind the trace list if the system error correction is switched off;
see also Calibration Overview.
Remote control:
[:SENSe<Ch>]:CORRection:STATe OBN | OFF
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......