![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 297](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759297.webp)
R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.135
E-6
Remote control:
[SENSe<Ch>:]SWEep:TPYE LINear
[SENSe<Ch/Tr>:]FUNCtion[:ON] "XFRequency:..."
Log. Frequency
In a
Log. Frequency
sweep the stimulus frequency is swept on a logarithmic scale over the continuous
frequency range. The frequency range (sweep range) is defined with the Stimulus settings. The sweep
points are calculated from the Span and the specified Number of Points (n > 1) with the condition that
the step width is constant on the logarithmic scale. The internal generator power can be set, if so
desired, in the Power Bandwidth Average submenu.
Log Frequency
sweeps are suitable for the analysis of a DUT over a large frequency range, e.g. over
several octaves. In a Cartesian diagram the measurement result is displayed as a trace over a
logarithmic frequency scale. The following example shows a
Log. Frequency
sweep with a stimulus
range between 50 MHz and 6 GHz, the forward transmission parameter S12 as measured quantity, and
a
dB Mag
scaled y-axis
Remote control:
[SENSe<Ch>:]SWEep:TPYE LOGarithmic
[SENSe<Ch/Tr>:]FUNCtion[:ON] "XFRequency:..."
Segmented Frequency
In a
Segmented Frequency
sweep the sweep range can be composed of several continuous, non-
overlapping frequency sub-ranges or single frequency points. The sub-ranges are termed sweep
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......