System Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.254
E-6
External Power Meters
Configures external power meters with their connection type and device address. The analyzer supports
external R&S NRP-Z sensors as power meters, to be connected via USB interface (the supported types
comprise NRP-Z21, NRP-Z11, NRP-Z22, NRP-Z23, NRP-Z24, NRP-Z51, NRP-Z52, NRP-Z55, NRP-
Z91).
Use of external power meters, examples
An external power meter can serve different purposes:
•
Extended measurement functionality: Each external power meter represents an additional
receive port. External generators increase the number of RF output signals of a DUT that the
analyzer can measure simultaneously. They can also provide accurate results for signals at
inaccurate or unknown frequencies. A typical example is a mixer measurement with an
unknown LO signal (and therefore unknown IF output frequency).
•
Power calibration: An external power meter can measure the exact signal power at an arbitrary
point in the test setup (reference plane) and thus provide the reference values for a power
calibration. A typical example is a source power calibration for an arbitrary analyzer port.
Configured external power meters can be selected as additional receivers in the
Port Configuration
and
in the power calibration dialogs.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......