![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 725](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759725.webp)
R&S
®
ZVA/ZVB/ZVT
SOURce<Ch>:...
1145.1084.12 6.263
E-6
SOURce<Ch>:...
This subsystem controls the frequency and power of the internal signal source, configures and starts
the source power calibration, and provides output port settings.
Note:
The SOURce<Ch>:POWer... subsystem comprises port-specific and general settings. Port-
specific settings are valid for the port specified by the numeric suffix
<Pt>
(...
:POWER<Pt>:
...).
General settings are valid for all test ports of the analyzer; the port suffix is ignored. Refer to the
description of the individual commands for more information.
SOURce<Ch>:FREQuency<Pt>[:CW] <fixed frequency>
SOURce<Ch>:FREQuency<Pt>:FIXed <fixed frequency>
Defines the fixed (Continuous Wave, CW) frequency for all sweep types operating at fixed frequency
(power sweep, time sweep, CW mode sweep). The two command forms
SOUR:FREQ:CW
and
SOUR:FREQ:FIX
are equivalent.
SOURce<Ch>:FREQuency<Pt>:CW|FIXed
is equivalent to
[SENSe<Ch>:]FREQuency:CW|FIXed
. Source and receiver frequency are equal; the four commands
overwrite each other. With option ZVA-K4,
Frequency Conversion,
independent offset frequencies can
be defined for the source and receiver frequencies; see
SOURce<Ch>:FREQuency<Pt>:CONVersion:ARBitrary:IFRequency
and
[SENSe<Ch>:]FREQuency:CONVersion:ARBitrary
.
<Ch>
Channel number. If unspecified the numeric suffix is set to 1.
<Pt>
Test port number of the analyzer. This suffix is ignored because
the selected frequency applies to all source ports used in the
active channel.
<fixed_frequency>
Fixed stimulus and analyzer frequency.
Range [def. unit]
Depending on the instrument model [Hz]. The increment
(parameters UP or DOWN) is 0.1 kHz.
*RST value
1 GHz
SCPI, Command
Types
Confirmed, command or query
Example:
FUNC "XPOW:POW:A1"
Activate a power sweep and select the wave quantity a1
as
measured parameter for channel and trace no. 1
.
SOUR:FREQ:CW 100MHz
Set the CW frequency to 100 MHz.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......