![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 515](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759515.webp)
R&S
®
ZVA/ZVB/ZVT
CALCulate
1145.1084.12 6.53
E-6
CALCulate<Ch/Tr>:DLINe:STATe <Boolean>
Switches the horizontal line on or off.
<Ch/Tr>
Channel number used to identify the active trace.
<Boolean>
ON | OFF - Horizontal line on or off.
*RST value
OFF
SCPI, Command Types
Device-specific, command or query.
Example:
See CALC:DLIN.
CALCulate<Ch/Tr>:FILTer[:GATE]...
This subsystem defines the properties of the time gate optimize the time domain response (with option
ZVAB-K2,
Time Domain)
.
CALCulate<Ch/Tr>:FILTer[:GATE]:TIME:CENTer <numeric_value>
Defines the center time of the time gate.
<Ch/Tr>
Channel number used to identify the active trace
<numeric_value>
Center time of the time gate
Range [def. unit]
-99.8999999 s to +99.8999999 s. The increment (parameters
UP or DOWN) is 0.1 ns. [s]
*RST value
1.5E-009 s
SCPI, Command
Types
Confirmed, command or query
Example:
*RST; CALC:TRAN:TIME:STAT ON;
CALC:FILT:TIME:STAT ON
Reset the instrument and enable the time domain
representation and the time gate.
CALC:FILT:TIME:CENT 0; SPAN 5ns
S
et the center time to 0 ns and the time span to 5 ns.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......