![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 529](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759529.webp)
R&S
®
ZVA/ZVB/ZVT
CALCulate
1145.1084.12 6.67
E-6
CALCulate<Ch/Tr>:LIMit:LOWer:SHIFt <numeric_value>
Shifts all lower and upper limit line segments assigned to the active trace in vertical direction.
This command is identical with CALCulate<Ch/Tr>:LIMit:UPPer:SHIFt.
<Ch/Tr>
Channel number used to identify the active trace.
<numeric_value>
Response offset value for all limit line segments.
Range [def. unit]
Almost no restriction for limit segments; see Rules for Limit Line
Definition. [dB]
*RST value
–
SCPI, Command
Types
Device-specific, no query.
Example:
See CALCulate<Ch/Tr>:LIMit:LOWer:FEED.
CALCulate<Ch/Tr>:LIMit:LOWer:STATe <Boolean>
Switches the lower limit check on or off. Lower limit line segments are assigned even numbers; see
CALCulate<Ch/Tr>:LIMit:LOWer[:DATA]
. The command does not affect segments with odd
numbers.
Use
CALCulate<Ch/Tr>:LIMit:STATe
to switch on or off the entire limit check, including
upper and lower limit lines.
<Ch/Tr>
Channel number used to identify the active trace.
<Boolean>
ON | OFF - Limit check on or off.
*RST value
OFF
SCPI,
Command
Types
Confirmed, command or query.
Example:
CALC:LIM:LOW -10, 0, 0, -10
Define the following lower and (default) upper limit line segments:
CALC:LIM:LOW:STAT ON; CALC:LIM:FAIL?
Switch the limit check on and query the result.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......