![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 695](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759695.webp)
R&S
®
ZVA/ZVB/ZVT
SENSe
1145.1084.12 6.233
E-6
[SENSe<Ch>:]LO...
This subsystem controls the Local Oscillator (LO) amplifiers for the reference and measurement
channels.
Command tree:
[SENSe<Ch>:]LO...
:LOMeasure
:LOReference
[SENSe<Ch>:]LOMeasure<Pt>[:STATe] <Boolean>
Switches the Local Oscillator amplifier for the measurement channel b on or off.
<Ch>
Channel number. If unspecified the numeric suffix is set to 1.
<Pt>
Test port number of the analyzer.
<Boolean>
ON | OFF – Switches the LO oscillator on or off.
*RST value
ON
SCPI, Command
Types
Device-specific, command or query
Example:
LOM1 OFF; LOR1 OFF
Switch the local oscillators for the measurement channel b and the
reference channel a at port no. 1 off.
[SENSe<Ch>:]LOReference[:STATe] <Boolean>
Switches the Local Oscillator amplifier for the reference channel a on or off.
<Ch>
Channel number.
<Pt>
Test port number of the analyzer.
<Boolean>
ON | OFF – Switches the LO oscillator on or off.
*RST value
ON
SCPI, Command
Types
Device-specific, command or query
Example:
LOM1 OFF; LOR1 OFF
Switch the local oscillators for the measurement channel b and the
reference channel a at port no. 1 off.
[SENSe<Ch>:]LPORt...
This subsystem controls the frequency reference oscillator.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......