R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.185
E-6
•
Renormalization of the physical port impedances affects e.g. S-parameters and wave quantities
in
Power
representation.
•
Renormalization of the balanced port impedances affects all measured quantities
(Trace –
Measure)
that the analyzer provides for balanced ports.
The default reference impedance of a physical port is equal to the reference impedance of the
connector type assigned to the port (50 or 75
).
It can be defined as a complex value.
For balanced ports it is possible to define separate complex reference impedances for differential and
for common mode.
The default values for the balanced port reference impedances are derived from the default reference
impedance of the physical analyzer ports (Z
0
= 50
):
•
The default value for the differential mode is Z
0d
= 100 = 2*Z
0
.
•
The default value for the common mode is Z
0c
= 25 = Z
0
/2.
Remote control:
SENSe:PORT<phys_port>:ZREFerence <real> [,<imaginary]
SENSe:LPORT<log_port>:ZCOMmon <real> [,<imaginary]
SENSe:LPORT<log_port>:ZDIFferent <real> [,<imaginary]
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......