![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 666](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759666.webp)
SENSe
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.204
E-6
Example:
CORR:COLL:METH REFL1
Select a one-port normalization at port 1 with an open standard as
calibration type.
CORR:COLL:SAVE:DEF
Calculate a dummy system error correction for the normalization at port
1. The dummy system error correction provides the reflection tracking
error term 'SCORR3'.
CORR:DATA? 'SCORR3'
Query the dummy system error correction term. The response is a 1
(written as 1,0 for the real and imaginary part) for each sweep point (no
attenuation and no phase shift between the analyzer and the calibration
plane).
CORR:DATA 'SCORR3',<ASCII_data>
Replace the dummy system error correction term with your own
correction data, transferred in ASCII format.
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected
Calculates the system error correction data from the acquired measurement results
(
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected),
stores them and applies them to
the calibrated channel <Ch>. To avoid incompatibilities, older system error correction data is deleted
unless it has been transferred into a cal pool (
MMEMory:STORe:CORRection <Ch>,
'<file_name>').
<Ch>
Channel number of the calibrated channel.
*RST value
–
SCPI, Command
Types
Confirmed, no query
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as
calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply them to the
active channel.
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault
Generates a set of default system error correction data for the selected ports and calibration type. The
default data set corresponds to a test setup which does not introduce any systematic errors; none of the
measurement results acquired previously
(
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected)
is taken into account.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......