R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.167
E-6
Always select the balanced port configuration
before
defining port groups. When a new balanced
port is created, the analyzer deactivates
Simultaneous Measurement of Port Group
and deletes all
existing port groups.
Remote control:
SOURce<Ch>:GROup<group_no> <log_port1>,<log_port2>
SOURce<Ch>:GROup<group_no>:CLEar [ALL]
Harmonics
The
Harmonics
submenu
controls the measurement of arbitrary harmonics or ratios between harmonics
and the fundamental wave. Together with the
Mixer
measurement
and the arbitrary configuration of
generator and receiver frequencies
(Port Configuration),
the harmonics measurement is included in
option R&S ZVA-K4.
Harmonics Measurements
Harmonics are signals at an integer multiple of the fundamental frequency. The fundamental is the first
harmonic, the n
th
harmonic is n times the frequency of the fundamental. The production of harmonic
frequencies by an electronic system when a signal is applied at the input is known as harmonic
distortion.
The purpose of the
Harmonics
menu is to measure the harmonic distortion of a DUT. To this end the
source remains at the fundamental frequency whereas the receiver is set to n times the fundamental
frequency. Two different types of results are provided:
•
In the direct measurement, the n
th
harmonic of the stimulus signal is measured. The result is
displayed as a wave quantity but can be modified and converted using the functions in the
Trace
menu.
•
In the relative measurement, the n
th
harmonic of the stimulus signal is divided by 1
st
harmonic
(fundamental) received from the DUT. The result corresponds to the n
th
harmonic distortion
factor.
Harmonic distortion measurements can be performed irrespective of the sweep type: A frequency
sweep yields the harmonics as a function of the frequency, a power sweep yields the power-dependent
harmonics at fixed frequency.
•
2nd Harmonic Port 1 to Port 2...
activates the measurement of the second harmonic at port 2
with port 1 as source port.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......