Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.206
E-6
Measure Standards
The last dialog of the calibration wizard is used to perform the necessary measurements of standards
and to calculate the correction data.
Measure Standards
displays the list of measured standards compiled in the previous dialogs. The list is
complete: All standards must be measured to perform the selected calibrations.
Structure of the
Measured Standards
list
The list of measured standards has a tree structure.
•
The first level contains all physical ports where one-port
(reflection) measurements are
required and all physical port combinations where two-port
(transmission) measurements
must be performed.
•
The second level contains check boxes for the standards to be measured at each port or port
combination.
•
For a sliding match, a third level contains check boxes for 7 different positions of the sliding
element. The sliding load must be measured at 3 to 7 different positions which should be
unequally spaced to avoid overlapping data points.
Double-click a physical port symbol to expand or collapse the list.
Checking one of the boxes in the list
causes the analyzer to stop the measurement in all channels
except the active one and measure the standard according to the active channel settings. The progress
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......