![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 392](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759392.webp)
Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.230
E-6
The circuit model for the load consists of capacitance C which is connected in parallel to an inductance
L and a resistance R, both connected in series.
•
R
is the constant resistive contribution. It is possible to select a special value
(Open
for
so
that the inductance coefficients are irrelevant,
Short
for 0
,
Match
for the reference impedance
of the current connector type) or set any resistance R.
•
The fringing capacitance
C
and the residual inductance
L
are both assumed to be frequency-
dependent and approximated by the first four terms of the Taylor series around f = 0 Hz.
Load parameters and standard types
Load parameters are used to describe all types of standards except a
Through,
a
Sliding Match,
aLineand an
Attenuation.
•
The
Through
standard is a through-connection between two ports with minimum loss which is
taken into account by the
Offset Parameters.
•
The
Sliding Match
is a one-port standard with variable load parameters (sliding load), so there
is no fixed load model.
•
The
Line
standard is a line of variable length with minimum loss which is taken into account by
the
Offset Parameters.
•
The
Attenuation
is a two-port standard which is fully matched in both directions (the reflection
factor at both ports is zero). No load and offset parameters need to be set.
Remote control:
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type>
Calibration Standard Types
The following table gives an overview of the different standards and their offset and load models:
Standard Type
Characteristics
Ideal Standard
(Dummy
Parameters)
Offset Model
Load
Model
Open Open
circuit
(one-port)
Short Short
circuit
(one-port)
0
Offset short
Short circuit with added electrical length
offset, for waveguide calibration (one-port)
0
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......