![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 512](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759512.webp)
CALCulate
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.50
E-6
Detailed Command Reference
CALCulate
This subsystem provides access to the results of a measurement.
The trace data is transferred in either ASCII or block data (REAL) format, depending on the
FORMat[:DATA] setting. If the block data format is used, it is recommended to select EOI as
receive terminator (
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator EOI
).
CALCulate<Ch/Tr>:DATA FDATa | SDATa | SCORr1 | ... | SCORr27
Reads the current response values of the active data trace or memory trace and reads or writes error
terms.
<Ch/Tr>
Channel number used to identify the active trace. If unspecified the
numeric suffix is set to 1.
Parameters
See list of parameters below.
Range [def.
unit]
The data format is parameter-dependent; see below. The unit is the
default unit of the measured parameter; see
CALCulate<Ch>:PARameter:SDEFine
.
*RST value
–
SCPI,
Command
Types
Confirmed, command (for error terms) or query.
Example:
*RST; SWE:POIN 20
Create a trace with 20 sweep points, making the created trace the
active trace of channel 1 (omitted optional mnemonic
SENSe1
).
CALC:DATA? FDAT
Query the 20 response values of the created trace. In the FDATa
setting, 20 comma-separated ASCII values are returned.
CALC:DATA:STIM?
Query the 20 stimulus values of the created trace. 20 comma-
separated ASCII values are returned.
The following parameters are related to trace data:
FDATa
Formatted trace data, according to the selected trace format (
CALCulate<Ch/Tr>:FORMat
). 1 value per
trace point for Cartesian diagrams, 2 values for polar diagrams.
SDATa
Unformatted trace data: Real and imaginary part of each measurement point. 2 values per trace point
irrespective of the selected trace format.
The following parameters denote the error terms generated during a calibration:
Error Term
Description
Receive Ports (S-
parameter)
SCORr1, ...
SCORr12
2-port error terms; see
[SENSe<Ch>:]CORRection:DATA.
1 and 2 (S11, S12, S21,
S22)
SCORr13
Directivity 3
(S33)
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......