ZVA/ZVB/ZVT
SENSe
1145.1084.12 6.191
E-6
Depending on the standard type, only a subset of the parameters may be used; see table below.
<Ch>
Channel number. This suffix is ignored because calibration kits are
channel-independent.
<std_type>
Standard type. For reflection standards, the first character denotes the
gender, e.g.:
FOPEN, MOPEN
: Open (f) or Open (m) standard.
The following reflection standards are supported:
MOPen, FOPen,
MSHort, FSHort, OSHort, MOSHort, FOSHort, MMTCh,
FFTCh, MREFLect, FREFLect
For transmission standards, the first two characters denotes the
genders on both ends, e.g.:
FFSNetwork, MFSNetwork, MMSNetwork:
Symm. network (ff),
symm. network (mf) or symm. network (mm) standard.
The symmetric network transmission standards quoted above
are supported.
See also the table of standard types below.
Parameter
list
String parameters to specify the configured standard
('<Conn_Name>','<Ckit_Name>','<Std_No>') and numeric parameters
defining its properties. See detailed description below.
The following transmission standards have only 8 parameters (until
<Z0>):
MMTHrough, MFTHrough, FFTHrough, MMLine, MFLine, FFLine
The following reflection and transmission standards have only 5
parameters (until <Max_Freq>):
MSMatch, FSMatch, MMATten, MFATten, FFATten
*RST value
–
SCPI,
Command
Types
Device-specific, command or query (see examples below)
Example:
CORR:CKIT:FOP 'N50 Ohm','Test Kit',...
Define the properties of the open (f) standard for a N 50
Ω
connector type
in cal kit 'Test Kit'. See also example for
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type>
.
CORR:CKIT:FOP? 'N50 Ohm'
Query the properties of the open (f) standard for a N 50
Ω
connector type
in the active cal kit.
CORR:CKIT:FOP? 'N50 Ohm','Test Kit'
Query the properties of the open (f) standard for a N 50
Ω
connector type
in cal kit 'Test Kit'.
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type>
'<Ckit_Name>','<Std_No>',<Min_Freq>,<Max_Freq>,<El_Length>,<Loss>,<C0> |
<L0>,<C1> | <L1>,<C2> | <L2>,<C3> | <L3>, OPEN | SHORt
Defines the parameters of a calibration standard <std_type> for a specified connector type
<conn_type>. A particular physical standard can be selected by specifying the name of the calibration
kit and its serial number. Depending on the standard type, only a subset of the parameters may be
used; see table below.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......