Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.176
E-6
•
RF
signal: Fixed port number (1), power sweep range (or fixed power, if the LO signal
is swept), CW frequency.
•
LO signal: Signal source (analyzer port or external generator), fixed power (or power
sweep range, if the RF signal is at a fixed power), CW frequency.
•
IF signal: Fixed port number (2), fixed frequency = (RF + LO) or |RF – LO|.
•
Time or CW mode sweep
•
For a
Time
or
CW Mode
sweep, the signal diagram is labeled as follows:
•
RF
signal: Fixed port number (1), fixed power, CW frequency.
•
LO signal: Signal source (analyzer port or external generator), fixed power, CW
frequency.
•
IF signal: Fixed port number (2), fixed frequency = (RF + LO) or |RF – LO|.
Set Power
The
Set Power
dialog defines the power of the RF and the LO signals.