ZVA/ZVB/ZVT
Frequency-Converting DUTs
1145.1084.12 9.13
E-4
Frequency-Converting DUTs
The analyzer provides a wide range of measurement modes for frequency-converting DUTs.
Frequency-converting measurements require option R&S ZVA-K4
(Frequency Conversion
, for R&S
ZVA and R&S ZVT network analyzers) or option R&S ZVB-K3
(Mixer and Harmonics
, for R&S ZVB
network analyzers).
Harmonics Measurements
In the
Harmonic Distortion
menu you can measure arbitrary harmonics or ratios between harmonics and
the fundamental wave (harmonic distortion factor).
To measure a harmonic of arbitrary order n...
1. Reset the analyzer to ensure that the following operating mode is set:
Sweep Type: Lin.
Frequency, Trace - Meas.: S21, Trace - Format: dB Mag
.
2. Open
the
Channel - Stimulus
menu and reduce the
Stop
frequency to the maximum frequency
of your analyzer divided by n where n is the order of the harmonic you want to measure.
3. Click
Channel - Mode - Harmonics - More Harmonics.
In the dialog opened select the
Harmonic
Order
you want to measure.
4. Adjust
the
Source
port and the receive port
(Harmonics Measured at)
to your test setup (for a
DUT connected between port 1 and 2 of the analyzer you can keep the default settings).
5.
Close
the
More Harmonics
dialog to enable the harmonics measurement and observe the
changed default trace
Trc1.
The trace shows the n
th
harmonic as a wave quantity b
2
.
The measured quantity is changed
automatically when the harmonics measurement is enabled.
Use 2nd Harmonic Port1 to Port2 and 3rd Harmonic Port1 to Port2 to activate the measurement
of the second and third harmonics with a single menu command.
To obtain a harmonic distortion factor...
1. Reset the analyzer to ensure that the following operating mode is set:
Sweep Type: Lin.
Frequency, Trace - Meas.: S21, Trace - Format: dB Mag
.
2. Open
the
Channel - Stimulus
menu and reduce the
Stop
frequency to the maximum frequency
of your analyzer divided by n where n is the order of the harmonic distortion factor you want to
measure.
3. Click
Channel - Mode - Harmonics - More Harmonics.
In the dialog opened select the
Harmonic
Order
you want to measure.
4. Adjust
the
Source
port and the receive port
(Harmonics Measured at)
to your test setup (for a
DUT connected between port 1 and 2 of the analyzer you can keep the default settings).
5.
Close
the
More Harmonics
dialog to enable the harmonics measurement
.
6. Click
Relative Measurement
and observe the new mathematical trace
Trc2.
The trace belongs to a new channel
Ch2
and shows the n
th
harmonic distortion factor as a wave
quantity b
2
.
With the default trace format the result is displayed in dB units.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......