![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 587](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759587.webp)
R&S
®
ZVA/ZVB/ZVT
CALCulate
1145.1084.12 6.125
E-6
<no>
Number of capacitance in circuit model. If unspecified the numeric
suffix is set to 1.
Range [def. unit]
1 to 2 [–] The number of capacitances depends on the selected
circuit model.
Parameters
Possible circuit models (character data); see circuit models.
*RST value
– (see command type description below)
<numeric_value>
Capacitance C<no> for the specified circuit model.
Range [def. unit]
–1mF to 1 mF [F]. The increment is 1 fF (10
–15
F).
*RST value
1 pF (10
–12
F)
SCPI, Command
Types
Device-specific, command or query. In the query form, the
<numeric_value> parameter must be omitted. The command
returns the capacitance value for the specified circuit model.
Example:
*RST; CALC:TRAN:VNET:SEND:DEEM:PAR:C2? CSC
Query the default capacitance C2 for the
Serial C, shunt C
circuit
model. The response is 1E-012 (1 pF).
CALC:TRAN:VNET:SEND:DEEM:PAR:C2 CSC, 2.2E-12
Increase the capacitance to 2.2 pF.
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>:PARa
meters:L<no> CSL | LSC | LSL | SLC | SCL | SLL,<numeric_value>
Specifies the inductance values L1, L2, L3 in the different circuit models for single ended port
deembedding.
<Ch>
Channel number.
<Ph_pt>
Physical port number
<no>
Number of inductance in circuit model. If unspecified the numeric
suffix is set to 1.
Range [def. unit]
1 to 3 [–] The number of inductances depends on the selected
circuit model.
Parameters
Possible circuit models (character data); see circuit models.
*RST value
– (see command type description below)
<numeric_value>
Inductance L<no> for the specified circuit model.
Range [def. unit]
–1H to 1 H [H]. The increment is 1 pH (10
–12
H).
*RST value
1 nH (10
–9
H)
SCPI, Command
Types
Device-specific, command or query. In the query form, the
<numeric_value> parameter must be omitted. The command
returns the inductance value for the specified circuit model.
Example:
*RST; CALC:TRAN:VNET:SEND:DEEM:PAR:L1? SLL
Query the default inductance L1 for the
Shunt L, serial L
circuit
model. The response is 1E-009 (1 nH).
CALC:TRAN:VNET:SEND:DEEM:PAR:L1 SLL, 2.2E-9
Increase the inductance to 2.2 nH.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......