![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 650](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759650.webp)
SENSe
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.188
E-6
[SENSe<Ch>:]AVERage[:STATe] <Boolean>
Enable or disable the sweep average.
<Ch>
Channel number.
<Boolean>
ON | OFF - Enables or disables the automatic calculation of the sweep
average over the specified number of sweeps
([SENSe<Ch>:]AVERage:COUNt).
*RST value
ON
SCPI,
Command
Types
Confirmed, command or query
Example:
AVER:COUN 15
Set the average factor for channel 1 to 15.
AVER ON
Enable the sweep average over the defined number of sweeps.
[SENSe<Ch>:]BANDwidth...
This subsystem sets the bandwidth of the IF measurement filter (resolution bandwidth).The forms
BANDwidth
and
BWIDth
are equivalent.
[SENSe<Ch>:]BANDwidth|BWIDth[:RESolution] <numeric_value>
Defines the resolution bandwidth of the analyzer
(Meas. Bandwidth)
.
<Ch>
Channel number 1 to 9999. If unspecified the numeric suffix is set
to 1.
<numeric_value>
Resolution bandwidth
Range [def. unit]
1 Hz to 5 MHz [Hz]. UP and DOWN increment/decrement the
bandwidth in 1-2-5 steps for each decade. The analyzer rounds up
any entered value between these steps and rounds down values
exceeding the maximum bandwidth.
*RST value
10 kHz
SCPI, Command
Types
Confirmed, command or query
Example:
BAND 1.1
Set a resolution bandwidth of approx. 1.1 Hz for channel 1.
BAND?
The analyzer returns the rounded bandwidth of 2 Hz.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......