R&S
®
ZVA/ZVB/ZVT
SENSe
1145.1084.12 6.199
E-6
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
'<file_name>',<analyzer_port_no>,<cal_unit_port_no>{,<analyzer_port_no>,<cal
_unit_port_no>}
Selects and initiates a one-port, two-port, three-port or four-port automatic calibration
at arbitrary
analyzer and calibration unit ports.
Use the simplified command
[SENSe<Ch>:]CORRection:COLLect:AUTO
if you only connect
matching pairs of analyzer and calibration unit ports.
<Ch>
Channel number of the calibrated channel.
'<file_name>'
Name and (possibly) directory of the cal kit file to be used for
the automatic calibration (string parameter):
If nothing is specified (empty string ' '), the
default cal kit file stored in the calibration unit is used.
This file should be used to operate the calibration unit
like in manual control; specifying another cal kit file is
primarily for service purposes.
A NWA cal kit file name *.calkit without path
denotes a specific cal kit file stored in the Calibration
Unit.
A NWA cal kit file name *.calkit with path
denotes a specific cal kit file stored in an arbitrary
directory.
<analyzer_port_no>
Port numbers of the analyzer. For an n-port automatic
calibration (n = 1 to 4), n arbitrary (not necessarily consecutive)
port numbers must be specified.
<cal_unit__port_no>
Port numbers of the cal unit. For an n-port automatic calibration
(n = 1 to 4), n arbitrary (not necessarily consecutive) port
numbers must be specified. It is possible to combine arbitrary
(not necessarily matching) pairs of analyzer and cal unit ports.
*RST value
–
SCPI, Command
Types
Device-specific, no query
Example:
CORR:COLL:AUTO:PORT ' ', 1, 2, 2, 4, 4, 1
Perform an automatic 3-port TOSM calibration at the analyzer
test ports 1, 2, and 4 using the calibration unit's default
calibration kit file and ports 2, 4, and 1 of the cal unit.
[SENSe<Ch>:]CORRection:COLLect:CONNection<port_no> N50FEMALE |
N50MALE | N75FEMALE | N75MALE | PC7 | PC35FEMALE | PC35MALE |
PC292FEMALE | PC292MALE
Additional parameters (): UFEMALE1 | UMALE1 | UFEMALE2 | UMALE2 |
SMAFEMALE | SMAMALE
Selects a connector type at a specified port <port_no> and its gender.
<Ch>
Channel number of the calibrated channel.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......