![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 660](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759660.webp)
SENSe
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 6.198
E-6
<Ch>
Channel number of the calibrated channel.
Parameters
Standard types:
Through, Open, Short, Match, Symmetric Network
(NET), Attenuation (ATT), Reflect, Sliding Match (SLIDe), Line
.
<port_no>
Port numbers of the analyzer. For a transmission standard (through,
line, attenuation, symmetric network) the input and output port numbers
must be specified, for reflection standards, only one port number is
required.
*RST value
–
SCPI,
Command
Types
Device-specific, no query
Example:
See [SENSe<Ch>:]
CORRection:COLLect:SAVE:SELected
.
[SENSe<Ch>:]CORRection:COLLect:AUTO '<file_name>',<port_no>{,<port_no>}
Selects and initiates a one-port, two-port, three-port or four-port automatic calibration
at arbitrary
analyzer ports and matching calibration unit ports.
Use the extended command
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
if you wish
to connect arbitrary pairs of analyzer and calibration unit ports.
<Ch>
Channel number of the calibrated channel.
'<file_name>'
Name and (possibly) directory of the cal kit file to be used for the
automatic calibration (string parameter):
If nothing is specified (empty string ' '), the default cal kit file
stored in the calibration unit is used. This file is also used in
manual control.
A NWA cal kit file name *.calkit without path denotes a
specific cal kit file stored in the Calibration Unit.
A NWA cal kit file name *.calkit with path denotes a specific
cal kit file stored in an arbitrary directory.
<port_no>
Port numbers of the analyzer. For an n-port automatic calibration (n =
1 to 4), n arbitrary (not necessarily consecutive) port numbers must be
specified. However, the port numbers of the analyzer and the
calibration unit must match: Port 1 of the analyzer must be connected
to port 1 of the calibration unit etc.
*RST value
–
SCPI,
Command
Types
Device-specific, no query
Example:
CORR:COLL:AUTO ' ', 1, 2, 4
Perform an automatic 3-port TOSM calibration at test ports 1, 2, and 4
using the calibration unit's default calibration kit file and test ports 1, 2,
and 4 of the cal unit.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......