Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.202
E-6
A successful calibration will supersede the previous calibration, discarding all previous system
error correction data. To keep older correction data you can transfer them into a
Cal Pool
using the
Calibration Manager.
The system error correction data determined in a calibration procedure are stored on the analyzer. You
can read these correction data using the remote control command
[SENSe<Ch>:]CORRection:CDATa.
You can also replace the correction data of the analyzer by
your own correction data sets.
Select Connectors
The first dialog of the calibration wizard displays a table to select the connectors and calibration kits for
all calibrated physical ports.
The table contains the following rows:
•
Physical Port Number #
The ports (and therefore the number of table rows) are determined by the active calibration type
selected in the
Start Cal
submenu. For a user-defined calibration
(Start Cal
–
Other...)
all
measurement ports of the analyzer are displayed regardless of their port configuration.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......