![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 315](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759315.webp)
R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.153
E-6
define the reference impedances at the ports.
The dialog contains the following additional controls:
•
If
Same Connector Type at All Ports
is active, the connector types at all ports (but not their
gender) are always adjusted to the current selection in the
Connector
column.
•
Measure "a" Waves at
selects the analyzer frequency for the measurement of the reference
waves a
n
. Depending on the measured quantity, the reference waves can be measured at the
receiver or at the source frequency.
•
Freq Conv Off
disables all frequency-converting modes, i.e. it resets all port frequencies to the
basic channel frequency and disables the
Harmonics
and the
Mixer
measurement (with option
R&S ZVA-K4).
Remote control:
[SENSe<Ch>:]CORRection:COLLect:CONNection:PORTs ALL | SINGle
[SENSe<Ch>:]FREQuency:CONVersion:AWReceiver[:STATe]
Logical Port
Defines the properties of the logical (balanced) ports defined in the
Balanced and Measured Ports
dialog. To display this section, at least one balanced port configuration must be defined and
Logical
Ports – Ref Impedance
must be selected in the
Displayed Columns
dialog.
•
# is the port number of the logical port. This number is independent of the physical port
numbers and of the ports that are combined to form a balanced port.
•
Ref Impedance
shows the complex reference impedances of the ports.
Reference impedance settings
The default reference impedance for a physical port is equal to the reference impedance of the
connector type assigned to the port but can be defined as an arbitrary complex value (renormalization
of port impedances). By changing the reference impedance, it is possible to convert the measured
values at 50 (75 ) into values at arbitrary port impedances.
For balanced ports it is possible to define separate complex reference impedances for differential and
for common mode.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......