Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.190
E-6
Remote
control:
CALCulate<Ch>:TRANsform:VNETwork:BALanced...
MMEMory:LOAD:VNETworks<Ch>:BALanced:DEEMbedding<Log_pt>
'<file_name>', PMAIn, PSECondary
MMEMory:LOAD:VNETworks<Ch>:BALanced:EMBedding<Log_pt>
'<file_name>', PMAIn, PSECondary
Alternate
Activates or deactivates the operating mode where the analyzer performs only one partial measurement
per sweep.
•
In normal sweep mode
(Alternate
disabled), the analyzer completes the necessary sequence of
partial measurements at each sweep point and obtains the result (measurement point) before
proceeding to the next sweep point.
•
Alternate
reverses the order of partial measurements and sweeps: A partial measurement is
performed at all sweep points (partial sweep) before the hardware settings are changed and the
next partial measurement is carried out in an additional sweep.
Advantages of both sweep modes, alternative settings
If the settling time between adjacent frequency points is larger than the settling time between the partial
measurements, then the
Alternate
measurementis faster than a normal sweep so that a smaller sweep
times can be set. On the other hand, an
Alternate
measurement provides a result only during the last
partial sweep.
Use the
Alternate
mode to increase the accuracy of measurements on DUTs with long level
settling times (e.g. quartzes, SAW filters). To measure DUTs with short settling times and obtain a trace
from the beginning of the sweep, deactivate the
Alternate
mode.
As an alternative to activating the
Alternate
mode, it is possible to insert a
Meas. Delay
before each
partial measurement and thus improve the accuracy. The delay slows down the measurement.
Relation to trigger settings
In triggered measurements,
Alternate
has no effect if the triggered measurement sequence is identical
to a single sweep point. The following table shows how the analyzer performs a sweep comprising m
sweep points, assuming that each of them requires n partial measurements.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......