![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 317](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759317.webp)
R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.155
E-6
Reference impedance settings
The default reference impedance for a physical port is equal to the reference impedance of the
connector type assigned to the port but can be defined as an arbitrary complex value (renormalization
of port impedances). By changing the reference impedance, it is possible to convert the measured
values at 50 (75 ) into values at arbitrary port impedances.
To perform a system error correction, the physical port must be available as a source and receive
port. This condition is met if it is selected as a measured port in the Balanced and Measured
Ports dialog (Meas is on).
To perform a system error correction, it is necessary that:
•
the physical port is available as a source port
•
both the measurement and the reference channel at the port can be measured
These conditions are met with the
Port Configuration
settings listed below.
Section Parameter Value
Meas On
on
Source On
on
Perm off
Receiver LO
a
on
LO b
on
Remote control:
[SENSe<Ch>:]CORRection:COLLect:CONNection<port_no>
[SENSe<Ch>:]CORRection:COLLect:CONNection:PORTs ALL
|
SINGle
SENSe:PORT<phys_port>:ZREFerence <real>
[,<imaginary]
Source
Shows and/or defines the RF generator settings for each physical port of the analyzer and for each
external generator. No source settings are provided for external power meters.
Most of the settings in this section require option ZVA-K4, Arbitrary Generator and Receiver
Frequencies. This option is available for ZVA and ZVT network analyzers only.
•
RF Off
switches the RF signal source at the port off (box checked) or on.
Dependencies:
The
column is disabled (grayed) while the RF power is switched off globally for the active channel.
The settings are restored as soon as the RF power is switched on again. RF Off and Meas are
independent from each other: While the RF power is switched off, the port is not necessarily
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......