![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 703](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759703.webp)
R&S
®
ZVA/ZVB/ZVT
SENSe
1145.1084.12 6.241
E-6
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol <Boolean>
Qualifies whether or not the
Meas. Bandwidth
can be set independently for each sweep segment.
<Ch>
Channel number 1.
<Seg>
Sweep segment number.
<Boolean>
ON: The bandwidth can be set independently for each sweep segment.
OFF: The bandwidth in all sweep segments is equal to the bandwidth for
unsegmented sweeps set via
SENSe<Ch>:BWIDth[:RESolution]
.
*RST value
OFF. The parameter is automatically switched to ON when a bandwidth
is entered using
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]
.
SCPI,
Command
Types
Device-specific, command or query
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default
settings and thus 10 kHz measurement bandwidth.
SEGM:BWID 1 MHZ
Increase the resolution bandwidth to 1 MHz.
SEGM:BWID:CONT OFF
Couple the bandwidths in all segments and reset the bandwidth in
segment no. 1 to the initial value.
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect NORMal | HIGH
Defines the
Selectivity
in sweep segment no. <Seg>. At the same time, the command activates
separate selectivity setting in all sweep segments
(
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol:SELect ON
).
<Ch>
Channel number 1.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......