SENSe
1145.1084.12 6.201
E-6
[SENSe<Ch>:]CORRection:COLLect:DELete ['<cal_name>']
Deletes system error correction data generated and stored previously
.
<Ch>
Channel number of the calibrated channel.
'<cal_name>'
Name of the calibration (string parameter) defined together with
calibration type
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
. If nothing
is specified the analyzer deletes the last system error correction stored
by means of
[SENSe<Ch>:]CORRection:COLLect:SAVE
.
*RST value
–
SCPI,
Command
Types
Device-specific, no query
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as
calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE; DEL
Calculate the system error correction data and apply them to the active
channel, then delete the data.
[SENSe<Ch>:]CORRection:COLLect:METHod FRTRans | TOM | TRM | TRL | TNA
| TOSM | FOPort1 | FOPort2 | FOPort12 | FOPTport | ROPTport | REFL1 | REFL2 |
REFL12 | TPORt | ETOM | ETSM
Selects a one-port or two-port calibration type at ports 1/2.
Use the generalized command
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
to
select the calibration type for arbitrary analyzer ports or a multiport calibration type.
<Ch>
Channel number of the calibrated channel.
Parameters
Calibration types:
TOM, TRM, TRL, TNA, TOSM, Full One Port, One
Path Two Port, Normalization
(REFL1, REFL2 and REFL12 for one-
port, TPORT for two-port)
, TOM Enhanced, TSM Enhanced
.
The numbers in the parameters denote the analyzer ports. Parameters
for two-port calibration types contain no numbers because the
command is only valid for ports 1 and 2.
*RST value
–
SCPI,
Command
Types
Confirmed (with device-specific calibration types), command or query. If
several calibration types are assigned to channel <Ch>, the query
returns a list of all calibration types.
Example:
See
[SENSe<Ch>:]
CORRection:COLLect:SAVE
.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......