R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.213
E-6
Port n Cal Power
The
Port n Cal Power
dialog defines the target power for the source power calibration. The settings are
port-specific.
The
Port n Cal Power
settings serve two different purposes:
•
Port Power Offset
defines a port-specific offset to the channel power. The actual output power
at the port is equal to the channel power P
b
(Channel – Stimulus – Power)
plus the
Port Power
Offset.
It is equal to the
Port Power Offset
(in dBm) if
0 dBm
is selected instead of
P
b
.
For
power sweep and the selection
P
b
,
the actual port power varies across the sweep, for other
configuration the port power is constant.
•
Cal Power Offset
specifies a gain (positive values) or an attenuation (negative values) in the
signal path between the source port and the calibrated reference plane. With a
Cal Power
Offset
of n dB, the target power at the reference plane is equal to the actual output power at the
port plus n dB. The
Cal Power Offset
has no impact on the source power.
Use of an amplifier in the signal path
Assume that a DUT requires a constant input power of +35 dBm, and that the measurement path
contains an amplifier with a 30 dB gain.
After a reset of the analyzer the channel power P
b
is zero dBm. With a
Port Power Offset
of +5 dB at the
calibrated source port and a
Cal Power Offset
of +30 dB, the source power calibration ensures that the
constant input power of +35 dBm is maintained across the entire sweep range. The actual output power
of the analyzer is +5 dBm.
Note that a power calibration with an appropriate
Cal Power Offset
will automatically prevent excess
input levels at the DUT.
Remote
control:
SOURce<Ch>:POWer[:LEVel][:IMMediate]:OFFSet <numeric_value>,
ONLY | CPAD
SOURce<Ch>:POWer<Pt>:CORRection:LEVel:OFFSet
Receiver Power Cal...
Opens the configuration dialog for a receiver power calibration. The receiver power calibration data is
acquired in a calibration sweep after appropriate selection of the calibrated wave quantity and the
source.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......