Channel Menu
R&S
®
ZVA/ZVB/ZVT
1145.1084.12 4.232
E-6
If a dispersive connector type (i.e. a waveguide; see
Offset Model
dialog) is assigned to a test
port that is related to a particular quantity, then the phase of the quantity is corrected using both
the (constant) specified length offset parameter and the dispersion effects due to the
connector.
Offset parameters; Application and effect
Offset parameters can be particularly useful if the DUT is connected to a transmission line that
cannot be taken into account in a calibration, e.g. because it is fixed to the DUT and cannot be
removed.
An offset is assigned to a particular port and affects the phase of all measurement parameters
related to this port. An offset for port 1 affects the S-parameters S
11
, S
21
, S
12
, S
31
..., the Z-
parameters Z
11
, Z
21
,Z
12
, Z
31
etc.
To account for the propagation in both directions, the phase shift of a reflection
parameter due to a given length offset is twice the phase shift of a transmission parameter. If,
at a frequency of 300 MHz, the electrical length is increased by 250 mm ( /4), then the phase
of S
21
increases by 90 deg, whereas the phase of S
11
increases by 180 deg.
If the trace is converted to the
Delay
format, changing the offset parameters simply shifts the
whole trace in vertical direction. The sign of the phase shift is determined as follows:
•
A positive offset parameter causes a positive phase shift of the measured parameter
and therefore reduces the calculated (residual) delay.
•
A negative offset parameter causes a negative phase shift of the measured parameter
and therefore increases the calculated (residual) delay.
Offset parameters for balanced ports
The functions in the
Offset
menu can be used for balanced port configurations:
•
If a balanced port configuration is active the logical and physical ports are shown in the
Electrical Length, Mechanical Length
and
Delay
dialogs.
•
Offset parameters must be assigned to both physical ports of a logical port.
•
Auto Length
corrects the length offset of both physical ports of a logical port by the
same amount.
The length offset can be defined separately for each port.
•
Reset Offsets
restores the default values for all length offsets, i.e. it resets all values to zero.
•
Electrical Length
defines electrical length offsets.
•
Mechanical Length
defines mechanical length offsets.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......