![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 579](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759579.webp)
R&S
®
ZVA/ZVB/ZVT
CALCulate
1145.1084.12 6.117
E-6
SCPI,
Command
Types
Device-specific, command or query
Example:
*RST; CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable the
time domain transformation for the default trace.
CALC:TRAN:TIME:XAX DIST
Convert the x-axis scaling to distance units.
CALCulate_TRANsform_VNETworks
CALCulate<Ch>:TRANsform:VNETworks...
This subsystem defines the circuit models for single ended and balanced port (de-)embedding and
activates the (de-)embedding function.
The circuit models are referenced by means of predefined character data parameters. They are
different for single ended and balanced port (de-)embedding.
Circuit models for balanced port embedding/deembedding:
Parameter Circuit
model
Capacitances Resistances Inductances
FIMPort File
import, no
circuit
model
– –
–
STSL Serial
Touchston
e (.s2p)
data,
shunt L
– R1
L1
STSC Serial
Touchston
e (.s2p)
data,
shunt C
C1 R1
–
SLST Shunt
L,
serial
Touchston
e (.s2p)
data
– R1
L1
SCST Shunt
C,
serial
Touchston
e (.s2p)
data
C1 R1
–
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......