Impact of offset model parameters
The offset model parameters are used for the calculation of the S-parameters for the calibration
standards associated with the connector type, provided that they are derived from a circuit model
(Add/Modify Standard
dialog).
•
For TEM-type lines, the relative permittivity
r
is needed for the conversion of and ZVR-type
Loss
(in units of dB/sqrt(GHz)) into an Agilent-type
Offset Loss
(in units of G /s) and vice versa
(see
Offset
and
Load
parameters). The
Electrical Length
and
Delay
values in the
Modify Offset
dialog are directly entered and therefore independent of
r
.
•
For waveguides, the low frequency cutoff frequency f
c
is important because no wave
propagation is possible at frequencies below f
c
. If a standard is measured in order to acquire
calibration data, the analyzer checks the low frequency cutoff. If the start frequency of the
sweep range is below f
c
, then the calibration wizard generates and error message.
The offset model parameters are not used except in the context of calibration. The offset
parameterdefinitions (see
Mechanical Length)
are based on independent
r
values.
Remote control:
[SENSe<Ch>:]CORRection:CONNection
Add or View / Modify Standard
Defines, displays or modifies the properties of the calibration standards in a particular calibration kit.
This dialog is opened from the
Add or View / Modify Calibration Kit
dialog
(Add Standard...
or
View /
Modify Standard...
buttons)
.
Depending on the title, some control elements may not be active.
For an overview of calibration standards and their properties refer to section Calibration standard
types below.
In its upper part the
Add Standard
or
View / Modify Standard
dialog
contains several controls to do the
following:
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......